Exercise 1
Propane, C3H8, and carbon dioxide, CO2, have the same nominal mass, 44. Calculate their exact masses (C = 12.0000; H = 1.0078; O = 15.9949) and the minimum resolution needed to separate their molecular ions.
C3H8: 3 × 12.0000 + 8 × 1.0078 = 44.0626. CO2: 12.0000 + 2 × 15.9949 = 43.9898. The difference is 0.0728, so R ≈ 44 / 0.0728 ≈ 600 is needed. That seems little, but a quadrupole works at nominal-mass resolution: it separates peaks one unit apart, which at m/z 44 means R ≈ 44. It would not distinguish them; a TOF or an Orbitrap would.
Exercise 2
In a time-of-flight analyser, an ion of m/z 100 reaches the detector after 20 µs. When will an ion of m/z 400 with the same charge arrive?
The flight time is proportional to √(m/z). Since 400 / 100 = 4 and √4 = 2, the ion at 400 will take twice as long: 40 µs.
Exercise 3
In an FT-ICR with a 14 T magnet, an ion of m/z 200 circles at a frequency of 1.07 MHz. At what frequency will an ion of m/z 400 circle in the same instrument? Why can this frequency be measured so precisely?
The frequency is f = zeB/(2πm): inversely proportional to m/z. The ion at 400 will circle at half the frequency, 0.535 MHz. Frequencies are measured very precisely because the ions make hundreds of thousands of turns during the measurement, and the longer a periodic signal is recorded, the better its frequency is known. It is the same reason NMR measures frequencies so precisely.