The IR spectrum: how it is recorded and read
An IR spectrum plots the radiation that passes through the sample against wavenumber. Each absorption appears as a downward band. The bands above 1500 cm−1 identify the functional groups; those below make up the fingerprint region, which is characteristic of each compound.
How it is recorded: the FT-IR spectrometer
The first IR spectrometers were dispersive: a prism or a diffraction grating spread out the radiation and the instrument scanned the spectrum one wavenumber at a time, which took several minutes. Today almost all of them are Fourier transform spectrometers (FT-IR), which measure all frequencies at once with a Michelson interferometer.

The beam splitter sends half of the radiation to a fixed mirror and the other half to a moving mirror. When the two beams recombine they interfere, and the result depends on the path difference between them and on each wavelength. As the mirror moves, the detector records the total intensity against the mirror position: the interferogram. A Fourier transform converts it into the spectrum, just as the FID is converted in NMR (see The response of the nuclei: FID and Fourier transform).
A scan takes about a second, so dozens of scans can be added to improve the signal-to-noise ratio. Before measuring the sample, a background spectrum is recorded without it, which the software subtracts afterwards to remove the absorptions of CO2 and water vapour in the air.
Sample preparation
Glass and water absorb strongly in the IR, so they cannot be used as a container or as a solvent. The usual techniques are:
- Liquid film: a drop of the neat liquid between two NaCl or KBr plates, which are transparent in the IR (and dissolve in water: the sample must be dry).
- KBr pellet: the solid is ground with powdered KBr and pressed into a transparent disc.
- Nujol mull: the solid is dispersed in a paraffin oil. Nujol adds its own C–H bands (2920, 2850, 1460 and 1375 cm−1), which must be ignored.
- Solution in CCl4 or CHCl3, in an NaCl cell. This is the technique that allows isolated molecules to be measured, without hydrogen bonds between them.
- ATR (attenuated total reflectance): the sample, solid or liquid, is pressed onto a diamond or zinc selenide crystal, and the radiation probes it a few microns below the surface. It needs no preparation and is now the most widely used technique.

How the spectrum is plotted
By tradition, the IR spectrum is drawn with the wavenumber decreasing from left to right, from 4000 to 400 cm−1, and with the transmittance (the percentage of radiation that passes through the sample) on the vertical axis. That is why the bands point downwards: a band that reaches 10 % transmittance is a strong absorption. Sometimes the absorbance, A = −log T, is plotted instead; it is proportional to the concentration (Beer–Lambert law) and gives upward peaks.


Bands are described by their position and appearance. Intensity is given with the abbreviations s (strong), m (medium) and w (weak), and shape as broad or sharp. The spectrum of butanone would thus be summarized as “2982 (m), 1717 (s), 1365 (m), 1175 (m) cm−1”.
The regions of the spectrum
Functional group bands fall in fairly well-defined regions, which follow the rules of Hooke's law:

- 4000–2500 cm−1: stretching of bonds to hydrogen, O–H, N–H and C–H.
- 2500–2000 cm−1: triple bonds (C≡C, C≡N) and cumulated double bonds (C=C=C, N=C=O). This region is almost always empty, so any band in it stands out.
- 2000–1500 cm−1: double bonds, above all C=O and C=C, and N–H bending.
- 1500–400 cm−1: the fingerprint region. This is where single-bond stretches (C–C, C–O, C–N, C–halogen) and bends appear, all strongly coupled to each other. The pattern is as characteristic of each compound as a fingerprint: two substances with the same fingerprint are almost certainly the same. It is used for comparison with reference spectra, for example those in the SDBS database.
How to read an IR spectrum
There is no need to try to assign every band: a handful of questions extracts almost all the useful information. A sensible order is:
- Is there a strong band between 1650 and 1850 cm−1? Then there is a carbonyl, and its position tells what kind.
- Are there bands between 3200 and 3600 cm−1? A broad band indicates O–H; one or two narrower ones, N–H. A very broad absorption from 3300 to 2500 cm−1 together with a carbonyl indicates a carboxylic acid.
- Where are the C–H stretches? Below 3000 cm−1 they belong to sp3 carbons; above, to sp2 carbons (alkenes, aromatics), and near 3300, to terminal alkynes.
- Is there anything between 2000 and 2300 cm−1? A nitrile or an alkyne.
- Are there bands near 1600 and 1500 cm−1 and below 900? Probably an aromatic ring.
The absence of a band is as informative as its presence: a spectrum with nothing between 1650 and 1850 cm−1 rules out any carbonyl, and one with nothing above 3100 rules out OH and NH groups.