Charged, flying molecules: the mass spectrometer
A mass spectrometer turns molecules into ions, separates them by their mass-to-charge ratio (m/z) and counts how many arrive at each value. The result, the mass spectrum, gives the mass of the molecule and, through its fragments, clues to its structure. Only a tiny amount of sample is needed, of the order of micrograms or even less.
Charged molecules in flight
Neutral molecules cannot be "weighed" one by one. But once they carry an electric charge, they can be accelerated and deflected by electric and magnetic fields, and their path depends on their mass. That is the idea behind mass spectrometry, and every instrument has the same three parts:
- The ion source, where the sample molecules are turned into ions.
- The analyser, which separates the ions by m/z.
- The detector, which counts the ions arriving at each m/z value.
The ions travel all the way in a high vacuum. If air molecules remained, the ions would collide with them before reaching the detector.

In the classic magnetic-sector design, the ions are accelerated by a potential difference and enter a magnetic field that bends their path. Light ions are deflected a lot and heavy ones only a little. For a given field, only ions of one m/z value follow the path to the detector. By slowly varying the field, all of them arrive, one after another. Other analysers exist today; we will see them in section 3.
What m/z means
The analyser does not separate ions by their mass, but by their mass divided by their charge, m/z, where m is expressed in atomic mass units (u, or daltons, Da) and z is the number of charges. In most spectra of small organic molecules the ions carry a single charge (z = 1), and then m/z equals the mass of the ion. With proteins and other large molecules, however, ions with many charges are common, as we will see in the ionisation section.
How to read a mass spectrum
The spectrum is drawn as a bar chart: m/z on the horizontal axis and, on the vertical axis, the abundance of each ion as a percentage of the most abundant one. Here is the spectrum of toluene, C6H5CH3, obtained by electron ionisation:

- The molecular ion, M+•, is the whole molecule minus one electron. It appears at m/z 92 and gives the molecular mass: it is the most valuable piece of information in the spectrum. Its symbol reminds us that it is both a cation (+) and a radical (•), because it has lost a single electron.
- The base peak is the most intense one, set at 100 %. In toluene it is at m/z 91: the molecular ion loses a hydrogen and forms a very stable cation, the tropylium ion, C7H7+. The base peak need not be the molecular ion.
- The other peaks, at lower m/z, are fragments: charged pieces into which the molecular ion has broken. Their pattern is characteristic of each compound, like a fingerprint.
- The small peak at m/z 93, M+1, comes from molecules containing one 13C atom. We will use it in the isotopes section.
Coupled to chromatography
In practice, the mass spectrometer is almost always coupled to a separation technique: gas chromatography (GC-MS) for volatile compounds or liquid chromatography (LC-MS) for polar and larger ones. The components of a mixture leave the column one after another and enter the spectrometer, which records the spectrum of each. By comparing these spectra with libraries of hundreds of thousands of compounds, substances are identified in complex mixtures: drugs in clinical and forensic analysis, pollutants in water, aroma components…
▶ Mass spectrometry
First of all, read the molecular ion in a mass spectrum: it gives the molecular mass. Then the isotopes help to deduce the formula, and the fragments give clues to the structure.