Ionization methods

A molecule must be charged before it can be analysed. Electron ionisation does it violently: it knocks out an electron and the molecule fragments heavily. Soft methods (chemical ionisation, electrospray, MALDI) add a proton or a cation with little energy, and almost the whole spectrum is the ion of the intact molecule, [M+H]+ or [M+Na]+.

Electron ionisation (EI)

This is the classic method. The sample, in the gas phase, passes through a beam of electrons accelerated to 70 eV. When one of them passes close to a molecule, it knocks out an electron:

M + e− → M+• + 2 e−

The result is the molecular ion, M+•, a radical cation. Ionising an organic molecule takes only 8–11 eV, so the molecular ion forms with a lot of excess energy, and a good part of it breaks into fragments. This has an advantage and a drawback:

  • Advantage: the fragment pattern is very informative and reproducible. Since every laboratory uses 70 eV, spectra can be compared with large libraries.
  • Drawback: sometimes the molecular ion fragments almost completely and is barely visible, as happens with many alcohols and highly branched hydrocarbons. The most important piece of data, the molecular mass, may be missing.

Moreover, EI requires a sample that can be vaporised without decomposing, so it is not suitable for large, polar or fragile molecules.

Chemical ionisation (CI)

Besides the sample, a reagent gas is introduced into the source in large excess (methane, isobutane or ammonia). The electron beam ionises mostly the gas, and the gas ions, such as CH5+ from methane, transfer a proton to the sample molecules:

M + CH5+ → [M+H]+ + CH4

Proton transfer leaves very little excess energy. The [M+H]+ ion, one unit above the molecular mass, is intense and there are few fragments. It is a good complement to EI when the molecular ion is not seen.

Diagram of chemical ionisation: in the chamber, the electron beam from the filament ionises methane, which forms CH₅⁺; this transfers a proton to the analyte M and the [M+H]⁺ ion leaves for the analyser
Chemical ionisation with methane: first the reagent gas is ionised (①) and then it protonates the sample (②).

Electrospray (ESI)

The sample, in solution, leaves through a capillary held at a high voltage and forms a spray of charged droplets. The solvent evaporates, the droplets shrink and eventually release gas-phase ions of the molecule. It is a very soft method: mostly ions of the intact molecule with an added proton or cation are seen, [M+H]+, [M+Na]+ or [M+K]+ in positive mode, or [M−H]− in negative mode, and sometimes clusters such as [2M+Na]+.

Diagram of electrospray: the solution leaves a metal capillary at +3–5 kV, forming a Taylor cone and a spray of charged droplets; as the solvent evaporates, the droplets shrink and split until they release gas-phase MH+ ions, which enter the spectrometer
Electrospray: from solution to gas-phase ions, at atmospheric pressure.

Let us look at paracetamol (C8H9NO2, M = 151):

Schematic mass spectra of paracetamol: with electron ionisation, molecular ion at 151 and base peak at 109; with electrospray, only ions of the intact molecule at 152, 174 and 325
Paracetamol with electron ionisation and with electrospray (schematic).
  • With EI the molecular ion is seen at m/z 151, but the base peak is at 109: the molecular ion loses a molecule of ketene (CH2=C=O, 42 u), leaving the 4-aminophenol ion. There are other minor fragments.
  • With ESI there are no fragments: only [M+H]+ at 152, [M+Na]+ at 174 (152 + 22) and a little of the dimer [2M+Na]+ at 325. To deduce the molecular mass, the proton or the sodium must be subtracted.
Loss of ketene from the molecular ion of paracetamol: a hydrogen from the methyl of the acetyl group moves to the nitrogen and the N–C(O) bond breaks, leaving the 4-aminophenol radical cation, m/z 109, and neutral ketene, H2C=C=O, 42 u
How the molecular ion of paracetamol loses ketene (EI).

Electrospray couples naturally with liquid chromatography (LC-MS), and today it is the most widely used method in pharmaceutical chemistry, metabolomics and proteomics.

Multiply charged ions

Large molecules with many basic groups, such as proteins, pick up several protons at once in electrospray. A series of [M+nH]n+ ions forms, each with a different number of charges:

Simulated electrospray spectrum of myoglobin: an envelope of peaks from m/z 700 to 1400, each with a different number of charges, from 12+ to 24+
The charge envelope of a protein (simulated).

Since m/z = (M + n·1.007) / n, a protein of almost 17,000 u appears between m/z 700 and 1400, within the range of any analyser. The charge and the mass are calculated from two consecutive peaks. If p1 < p2, the charge of the second is n = (p1 − 1.007) / (p2 − p1). For example, with the peaks at 942.75 and 998.15: n = 941.74 / 55.40 = 17, and M = 17 × (998.15 − 1.007) = 16,951 u. Software performs this calculation, called deconvolution, automatically.

MALDI

In matrix-assisted laser desorption/ionisation (MALDI), the sample is mixed with a large amount of an organic matrix that absorbs laser light. A laser pulse vaporises the matrix, which carries the sample molecules with it and transfers a proton to them. It gives mostly singly charged ions, [M+H]+, even for molecules of hundreds of thousands of units, and is usually combined with a time-of-flight analyser. It is used for proteins, polymers and even to identify bacteria in hospitals.

Diagram of a MALDI-TOF: a UV laser pulse strikes the sample, mixed with a crystalline matrix on a plate at +20 kV; the matrix vaporises and protonates the sample; the MH+ ions are accelerated towards a grid and travel along a flight tube, where the light ones reach the detector first
MALDI coupled to a time-of-flight analyser (MALDI-TOF).

Soft ionisation methods for biomolecules earned John B. Fenn (electrospray) and Koichi Tanaka (laser desorption) the 2002 Nobel Prize in Chemistry.

MethodSampleMain ionsFragmentation
EIgas, volatile and stableM+•extensive
CIgas, volatile[M+H]+little
ESIsolution, polar[M+H]+, [M+Na]+, [M−H]−, [M+nH]n+almost none
MALDIsolid with matrix[M+H]+, [M+Na]+almost none

EI gives the molecule's fingerprint at the expense of the molecular ion. Soft methods give the molecular mass at the expense of structural information. That is why they are used together.